Ammonium measurement directly in the medium without sample preparation and sample transfer. Measurement of centrate and other process waters up to 2,000 mg/l NH4-N.
The ISE sensor AmmoLyt®Plus 700 IQ allows ammonium measurement directly in the process. Time-consuming steps such as sampling and preparation are eliminated. For stable measured values, the interference substance potassium is compensated simultaneously and can be displayed if needed. An energy-optimized operation by a suitable regulation of the blower aggregates is made possible by this ISE sensor. This will help effectively reduce operating costs of the system, which is amortized by the system in a very short time. Starting now, you will be able to reliably determine centrate and other process water with high ammonium load up to 2.000 mg/l NH4-N with the AmmoLyt®Plus!
- Calibration free, long-term stability
- No chemicals
- Short response times
- Measuring method: Electrochemical
- Appropriate Electrode: Reference electrode VARiON® Ref, Measuring electrode VARiON®Plus NO3, Compensation electrode VARiON®Plus Cl
- Measuring range/Resolution: NH4-N: 1 … 2,000 mg/l / 1 mg/l; 0.1 … 100 mg/l / 0.1 mg/l, NH4+: 1 … 2,580 mg/l / 1 mg/l; 0.1 … 129.0 mg/l / 0.1 mg/l
- Compensation Range: K+: 0.1 … 1,000 mg/l / 0.1 mg/l
- Measuring Accuracy in laboratory standard solutions: ± 5 % of measured value ± 0.2 mg/l in standard solutions
- Calibration Procedures: Matrix adjustment against any reference value, 2-point-calibration possible with multiple standard solution
- Working Life (typically): Reference electrode: 18 months, measuring and compensation electrode: 18 months (in typical application – municipal sewage plants)
- Temperature Measurement and Compensation: Integrated NTC thermistor, Range 32 °F … 104 °F (0 °C … +40 °C), Accuracy ±0.5 K, Resolution 0.1 K, t95 < 20 s
- Pressure Resistance: Maximum 0.2 bar (incl. SACIQ sensor connection cable, with installed electrodes)
- Ambient Conditions: Operating temperature: 32 °F … 104 °F (0 °C … +40 °C), storing temperature: 32 °F … 104 °F (0 °C … +40 °C)
- Electrical connections: 2-wire shield cable with quick fastener to sensor
- Electromagnetic Compatibility: EN 61326, Class B, FCC Class A; Intended for indispensable operation
- Certifications: CE
- Mechanical: Sensor body: V4A stainless steel 1.4571 Protective cup: POM, Temperature sensor: V4A stainless steel 1.4571 Protection rating: IP 68 (0.2 bar, with installed electrodes), Electrode connector: POM
- Weight: Approx. 1.48 lb (670 g, without electrode, without sensor connection cable)
- Warranty: AmmoLyt®Plus 700 IQ: 2 years, Electrodes: 1 year for defects of quality